Using the Interlock Key
Background
When the INTERLOCK function is set to ON,
tests are only allowed to start when both
Interlock pins on the signal I/O port are
shorted. Using the Interlock key will short the
INTERLOCK1 and INTERLOCK2 pins on the
signal I/O port.
See page 99 for the Signal I/O pin assignment.
tests are only allowed to start when both
Interlock pins on the signal I/O port are
shorted. Using the Interlock key will short the
INTERLOCK1 and INTERLOCK2 pins on the
signal I/O port.
See page 99 for the Signal I/O pin assignment.
Panel operation
1. Insert the Interlock
key into the SIGNAL
I/O port on the rear
panel.
I/O port on the rear
panel.
2. Set the INTERLOCK option to ON
in the Common Utility.
Page 76
Note
With INTERLOCK set to ON, the tester can now
only start a test when the Interlock key is
connected.
Set INTERLOCK to OFF to disable this feature.
only start a test when the Interlock key is
connected.