Fluke 1507 Insulation measuring device, 50 V, 100 V, 250 V, 500 V, 1000 V (+20 %, -0 %) 2427890 User Manual

Product codes
2427890
Page of 38
1507/1503 
Users Manual 
24 
Model 1503 
Output Voltage 
Display Range Resolution  Test 
Current 
Accuracy 
±
 (% of Rdg + Digits) 
0.1 to 20.00 M
Ω 
0.01 M
Ω 
20.0 to 200.0 M
Ω 
0.1 M
Ω 
500 V 
(0 % to + 20 %) 
200 to 500 M
Ω
 1 
M
Ω
 
1 mA @ 500 k
Ω 
±
 (2.0 % + 5) 
0.1 to 200.0 M
Ω 
0.1 M
Ω 
1000 V 
0 % to + 20 %) 
200 to 2000 M
Ω
 1 
M
Ω
 
1 mA @ 1 M
Ω 
±
 (2.0 % + 5) 
EN61557 Specification 
The following tables are a requirement for European labeling. 
Measurement Intrinsic 
Uncertainty Operating 
Uncertainty
1
 
Volts 
±
 (2.0 % + 3) 
30 % 
Earth-Bond Resistance 
±
 (1.5 % + 3) 
30 % 
Insulation Resistance 
Depends on test voltage and range. 
See Insulation Test specifications. 
30 % 
1.  This specification comes from the standard and indicates the maximum amount allowable by the standard.