Fluke Fluke 4325162 User Manual
Product codes
4325162
Earth/Ground Tester
Advanced Operation
17
c) Prolonged display test
By keeping the "DISPLAY MENU" button pressed during power on, the display test can
be prolonged for any length of time. Return to the standard operation mode by pressing
any button or turning of the central rotary switch.
By keeping the "DISPLAY MENU" button pressed during power on, the display test can
be prolonged for any length of time. Return to the standard operation mode by pressing
any button or turning of the central rotary switch.
d) Number of software version
By keeping the "SELECT" button pressed during the power on sequence, the software
version is indicated on the display. By pressing the "DISPLAY MENU" button a switch
over to the last calibration date is possible. This display sequence is terminated by
turning the central rotary switch or pressing the "START TEST" button.
By keeping the "SELECT" button pressed during the power on sequence, the software
version is indicated on the display. By pressing the "DISPLAY MENU" button a switch
over to the last calibration date is possible. This display sequence is terminated by
turning the central rotary switch or pressing the "START TEST" button.
Display format:
SOFTWARE-version:
X. X X
The measuring functions have two initial operational modes: the Control loop and the
Measuring loop.
Measuring loop.
Control Loop
After turning the function rotary switch, the voltage display mode is reached. Pushing
"DISPLAY MENU" now calls up the control loop. According to the selected measuring
function, different setting values can be displayed and changed in the control loop. The
"DISPLAY MENU" button switches between the different set values inside a continuous
loop. The "SELECT" button selects the setting to be changed. Pushing the "CHANGE
ITEM" button the instrument either switches between certain set values or increases the
decimal point selected with "SELECT" by 1.
"DISPLAY MENU" now calls up the control loop. According to the selected measuring
function, different setting values can be displayed and changed in the control loop. The
"DISPLAY MENU" button switches between the different set values inside a continuous
loop. The "SELECT" button selects the setting to be changed. Pushing the "CHANGE
ITEM" button the instrument either switches between certain set values or increases the
decimal point selected with "SELECT" by 1.
After parameter setting has been finished the next display can be called with "DISPLAY
MENU" or the measurement can be started with "START TEST".
MENU" or the measurement can be started with "START TEST".