Fluke Fluke 4325162 User Manual

Product codes
4325162
Page of 56
 
Earth/Ground Tester 
 
Export Stored Data to PC 
39 
4.  Display shows "C ON". In the "C ON" state the CODE function can be disabled by 
pressing the "CHANGE ITEM" key. The display then shows "C OFF". 
5.  If this display is acknowledged by pressing the "DISPLAY MENU" key, the user code 
and all changes of the limit values are erased. The original default values are 
restored into memory. 
6.  Now a new CODE number may be programmed and used for setting new 
parameters. 
Export Stored Data to PC 
Test data is automatically stored for all tests as a .csv file. Table 8 (continued on page 
40)
 is an example of the .csv file. 
To export data from the Tester to a PC: 
1.  Connect the USB cable from the Tester to the PC. 
2.  Use Windows Explorer on the PC to find new EGT drive in the Devices list.  
3.  Locate the Data.csv file on the EGT drive. 
4.  Use the standard PC tools to copy the file to a new location. 
Table 8. Sample .CSV File for Logged Data 
Measu
re
men
T
imestamp
 
Measu
re
men
Mode
 
Measu
re
men
Vo
ltag
e Um 
Measu
re
men
Fre
que
nc
y Fm 
In
terferen
c
Vo
ltag
e Ust 
15th Oct 2013  20:13:55 
3-pole R
E
 
48 V 
128 Hz 
0.0 V 
15th Oct 2013  20:15:55 
4-pole R
E
 
48 V 
128 Hz 
0.0 V 
15th Oct 2013  20:17:15 
3-pole Selective 
48 V 
128 Hz 
0.2 V 
15th Oct 2013  20:21:10 
4-pole Selective 
20 V 
111 Hz 
0.0 V 
15th Oct 2013  20:23:25 
2-pole AC Resistance 
48 V 
128 Hz 
0.2 V 
15th Oct 2013  20:24:48 
2-pole DC Resistance 
48 V 
NA  
0.2 V 
10th Nov 2013  20:24:48 
4-pole Re 
48 V 
111 Hz 
0.0 V 
10th Nov 2013  20:28:48 
4-pole Selective 
48 V 
128 Hz 
0.0 V