42 
Specifications 
Temperature Range 
 
Operating: 
0 °C to +35 °C (+32 °F to +95 °F) 
Storage: 
-30 °C to +60 °C (-22 °F to +140 °F) 
Temperature coefficient: 
±0.1 % of rdg / °C (below 18 °C and above 28 °C) 
Operating humidity: 
<95% RH non-condensing 
Operating altitude: 
2000 m 
Climatic Class: 
C1 (IEC 654-1) -5 °C to +45 °C, 5% to 95% RH 
Protection type  
Case: IP56
Battery door: 
IP40 
Electromagnetic compatibility: 
Complies with IEC61326-1: Portable 
Safety: 
Complies with IEC 61010-1: CAT None, Pollution Degree 2  
Measurement time: 
6 seconds, typical 
Maximum overload: 
250 V
rms
 (in association with misuse) 
Batteries: 
6 x 1.5v, AA LR6 Alkaline 
Battery life span: 
>3000 measurements, typical, RH + RE < 1 kOhm 
>6000 measurements, typical, RH + RE > 10 kOhm 
Dimensions: 
240 mm x 180 mm x 110 mm (9.5 in x 7.1 in x 4.4 in) 
Weight with batteries: 
1.52 kg (3.35 lb) 
Memory: 
Internal memory storage up to 1500 records 
accessible via USB port 
Interference Voltage DC + AC (U
ST
) Measurement 
Measuring method: 
fullwave rectification 
Measuring 
Range 
Display 
Range 
Resolution Frequency 
Range 
Accuracy 
1…50 V 
0.0…50 V 
0.1 V 
DC/AC 45…400 Hz sine 
± (5% of rdg +5 digit) 
 
Measuring sequence: 
approximately 4 measurements/s  
Internal resistance: 
approximately 1.5 M
Ω
 
Maximum overload: 
Urms = 250 V