Earth/Ground Tester 
 
Specifications 
47 
 
Measuring sequence: 
approximately 2 measurements/s  
Measuring time: 
typical 4 seconds includes reversal of polarity (2-pole or 
4-pole) 
Maximum interference voltage: 
≤3 V ac or dc, with higher voltages measurement will not 
be started 
Maximum inductivity: 
2 Henry  
Maximum overload: 
U
rms
 = 250 V  
Compensation of Lead Resistance (R
K
) 
Compensation of lead resistance (R
K
) can be switched on in functions R
E
 3pole, R
E
 3pole , 
R~, and R  2pole 
Formation of measured value: 
R
display
 = R
measured
 - R
compensated
 
*
 
Value of setpoint entry R
K
 = 0.000 
Ω
, variable from 0.000…29.99 
Ω
 by means 
of measuring adjustment.
 
Stakeless Ground Loop Measurement (
) 
Resolution Measurement 
range 
Accuracy  Operating 
Error 
0.001 to 0.1 
Ω
 0.020 
Ω
 to 199.9 
Ω
 
±
(7% rdg + 3 d) 
±
(10% rdg + 5 d) 
Measuring principle: Stakeless measurement of resistance in closed loops using two current 
clamps  
Measurement voltage: 
Um = 48 V ac (primary) 
Measurement frequency: 
128 Hz 
Noise current (I
ext
): max 
I
ext
 = 10 A (ac) (Ra < 20 
Ω
 max 
I
ext
 = 2 A (ac) (Ra > 20 
Ω
The information about stakeless ground loop measurements is only valid when used in 
conjunction with the recommended current clamps at the minimum distance specified.