Fluke Fluke 4325181 User Manual

Product codes
4325181
Page of 56
 
Earth/Ground Tester 
 
Advanced Operation 
25 
Press "START TEST" button.  
Now a fully automated test sequence of all relevant parameters like auxiliary earth 
electrode and probe and earth electrode resistance is implemented and finished with the 
display of the result R
E
.  
1.  Read out measured value R
E
2. Call 
R
S
 and R
H
 with "DISPLAY MENU". 
Remarks for the Setting of Earth Spikes 
Before setting the earth spikes for probe and auxiliary earth electrode make sure that the 
probe is set outside the potential gradient of earth electrode and auxiliary earth electrode. 
Such a condition is normally reached by allowing a distance of >20 m between the earth 
electrode and the earth spikes as well as to the earth spikes to each other. An accuracy 
test of the results is made by another measurement following repositioning of the 
auxiliary earth electrode or probe. If the value stays the same, the distance is sufficient. If 
the measured value changes, probe or auxiliary earth electrode must be repositioned 
until the measured value RE stays constant. 
Spikes wires should not run too close.  
3-pole Measurement with Longer Earth Electrode Connecting Leads 
1.  Use one of the accessory cable drums as earth electrode connecting lead. 
2.  Spool off cable completely and compensate line resistance as described in 
“Compensation of Earth Electrode Connecting Lead”.  
Time Average Measurement 
If there is a warning "MEASURED VALUE UNSTABLE" (see Table 5) after a test 
sequence, most likely it is caused by strong interference signals (for example, unsteady 
noise voltage). To get reliable values, the instrument offers the possibility of averaging 
over a longer period. 
1.  Select a fixed frequency (see Control Loop). 
2.  Keep the "START TEST" button pressed until the warning "MEASURED VALUE 
UNSTABLE" disappears. Maximum averaging time is approximately 1 minute.