Fluke Fluke 4325181 User Manual

Product codes
4325181
Page of 56
1625-2 
Users Manual 
44 
 
Measuring 
Range 
Display Range 
Resolution 
Accuracy 
Operating Error 
0.020 
Ω
…300 k
Ω
 
0.001 
Ω
...2.999 
Ω
 0.001 
Ω
 
 
 
3.00 
Ω
...29.99 
Ω
 0.01 
Ω
 
 
 
30.0 
Ω…
299.9 
Ω
 0.1 
Ω
 
± (2 % of rdg +2 digit) 
± (5% of rdg +5 digit) 
0.300 k
Ω…
2.999 k
Ω
 1 
Ω
 
3.00 k
Ω…
29.99 k
Ω
 10 
Ω
 
 
 
30.0 k
Ω…
299.9 k
Ω
 100 
Ω
 
 
 
 
Measuring time: 
typical 8 seconds with a fixed frequency 
 
30 seconds maximum with AFC and complete cycle of all 
measuring frequencies 
Additional error because of probe 
and auxiliary earth electrode 
resistance: 
digits
R
R
R
E
S
H
5
%
10
25
.
1
)
2000
(
6
+
×
×
Ω
+
 
Measuring error of RH and RS: 
typical 10 % of 
H
s
E
R
R
R
+
+
 
Maximum probe resistance: 
≤1 M 
Ω
 
Maximum auxiliary earth electrode 
resistance: 
≤1 M 
Ω
 
If after a measurement of probe, auxiliary earth electrode, and earthing resistance, a 
measurement error of higher than 30 % is assumed because of the influencing conditions, the 
display shows a warning symbol  and a notice that R
S
 or R
H
 are too high. 
 
R
H
 with Umeas = 48 V 
R
H
 with Umeas = 20 V 
Resolution 
<300 
Ω
 <250 
Ω
 1 
m
Ω
 
<6 k
Ω
 <2.5 
k
Ω
 10 
m
Ω
 
<60 k
Ω
 <25 
k
Ω
 100 
m
Ω
 
<600 k
Ω
 <250 
k
Ω
 1 
Ω