9  Fluke Corporation     1623-2 and 1625-2 GEO Earth Ground Testers
Compensation of lead resistance (R
K
)
Compensation of lead resistance (R
K
) can be switched on in functions R
E
 3-pole, R
E
 4-pole 
, R~ , and R  2-pole
Formation of measured value
R
display
 = R
measured
 - R
compensated
*
* Value of setpoint entry R
K
 = 0.000 
Ω
, variable from 0.000 to 29.99 
Ω
 by means of measuring adjustment.
Stakeless ground loop measurement (
)
Switch position
Resolution
Measuring range
Accuracy
Operating error
R
A
 4-pole    
0.001 
Ω to 0.1 Ω
0.02 
Ω to 199.9 Ω
± (7 % rdg + 3 d)
± (10 % rdg + 5 d)
Measuring principle: Stakeless measurement of resistance in closed loops using two current transformers 
Measuring voltage
Vm = 48 V ac (primary)
Measuring frequency
128 Hz
Noise current (I
EXT
)
Max. I
EXT
 = 10 A (ac) (R
A
 < 20 
Ω)
Max. I
EXT
 = 2 A (ac) (R
A
 > 20 
Ω)
Automatic range selection
The information regarding stakeless ground loop measurements is only valid when used in conjunction with the recommended current clamps at the minimum 
distance specified.