Figure 8. Measuring Polarization Index and Dielectric
Absorption Ratios
Using the Compare Function (Model
1507)
1507)
Use the Compare function to set a pass/fail compare level
for the insulation measurements. To use the Compare
function:
for the insulation measurements. To use the Compare
function:
1.
Press the
C
button to select the desired compare
value. You can choose from 100 k
Ω
, 200 k
Ω
, 500 k
Ω
,
1 M
Ω
, 2 M
Ω
, 5 M
Ω
, 10 M
Ω
, 20 M
Ω
, 50 M
Ω
, 100 M
Ω
,
200 M
Ω
, and 500 M
Ω
.
2.
Perform insulation tests as described earlier in this
manual.
manual.
3.
The green pass indicator will appear if the measured
value is greater than the selected value.
value is greater than the selected value.
4.
Press and hold the
C
button for 1 second to
disable the Compare function. The pass indicator will
turn off when you start a new test or choose a new
compare value.
turn off when you start a new test or choose a new
compare value.