Fluke 1503 Insulation measuring device, 500V, 1000 V (+20%, -0%) 2427883 User Manual

Product codes
2427883
Page of 38
 
Insulation Testers 
 
Specifications   
  
21 
IP Rating..............................................................IP40 
Altitude ................................................................Operating: 2000 m CAT IV 600 V, 3000 m CAT III 600 V 
Non Operating (Storage): 12,000 m 
Over-Range Capability ........................................110% of range 
Compliance to EN 61557 ....................................IEC61557-1, IEC61557-2, IEC61557-4, IEC61557-10 
Model 1503 Accessories .....................................TL224 Leads 
TP74 Probes 
Clips PN 1958654 (red) and PN 1958646 (black) 
Holster 
Model 1507 Accessories .....................................TL224 Leads 
TP74 Probes 
Clips PN 1958654 (red) and PN 1958646 (black) 
Holster 
Remote Probe 
AC/DC Voltage Measurement 
Accuracy 
Range Resolution 
50 Hz to 400 Hz 
±
 (% of Rdg + Digits) 
600.0 V 
0.1 V 
+ (2 % + 3) 
 
Input Impedance..................................................3 M
Ω
 (nominal), <100 pF 
Common Mode Rejection Ratio  
(1 k
Ω
 unbalanced)...............................................>60 dB at dc, 50 or 60 Hz 
Overload Protection.............................................600 V rms or dc