Earth/Ground Tester 
 
Operation 
17 
R
A
 4-Pole Selective Earth Resistance Measurement with Current Clamp 
The R
A
 4-pole Selective Earth Resistance Measurement with Current Clamp procedure is 
useful for the resistance measurement of different parallel sections of an earth/ground 
system. 
1.  Select function 4 POLE . See Figure 7. 
2.  Connect test leads. 
Connect terminals E/C1 and ES/P1 with the supplied safety test leads (1.5 m) to 
the earth electrode to be measured. Place two ground stakes in earth/dirt. 
Minimum distance between earth electrode (E/C1), probe (S/P2) and auxiliary 
earth (H/C2) should be a minimum 20 m. The  test lead eliminates the influence 
of the test leads. 
Connect stakes with 25 m and 50 m wires to H/C2 and S/P2 as shown. 
Connect current clamp with adapter cable as shown.  
3.  Push START
 indicates that measurement is in progress. For continuous 
measurement, continue to push the START button. 
4.   indicates completed measurement. The result is kept on display until a new 
measurement is started or the rotary switch is turned.