Curve 3: 
With increasing depth 
ρ
E is not decreasing: a strip conductor electrode is 
advisable. 
As measuring results are often distorted and corrupted, for example, by 
underground pieces of metal and underground aquifers, a second 
measurement, in which the stake axis is turned by an angle of 90 °, is 
always advisable (see graph below). 
 
H
S
E
ES
90°
I
 
edw022.eps 
Export Stored Data to PC 
Test data is automatically stored for all tests as a .csv file. Table 5 is an example of the 
.csv file. 
To export data from the Tester to a PC: 
1.  Connect the USB cable from the Tester to the PC. 
2.  Use Windows Explorer on the PC to find new EGT drive in the Devices list. 
3.  Locate the Data.csv file on the EGT drive. 
4.  Use the standard PC tools to copy the file to a new location. 
Table 5. Sample .CSV File for Logged Data 
Measurement Timestamp 
Measurement 
Mode 
Earth Ground 
Resistance R
E
 
Error 
Status 
15th Oct 2013  20:13:55 
3-pole R
E
 1.022 
Ω
 NA 
15th Oct 2013  20:13:55 
4-pole R
E
 1.022 
Ω
 NA 
15th Oct 2013  20:13:55 
3-pole Selective 
1.022 
Ω
 NA 
15th Oct 2013  20:13:55 
4-pole R
E
 NA 
Rh 
Limit 
Delete Stored Data 
To delete stored data in the Tester: 
1.  Connect the USB cable from the Tester to the PC. 
2.  Use Windows Explorer on the PC to find new EGT drive in the Devices list.  
3.  Locate the Data.csv file on the EGT drive. 
4.  Use the standard PC tools to delete the file from the EGT drive or move the file to a 
new location.  
This action removes all stored date from the Tester.