Error Codes 
Various error conditions are detected by the tester and are indicated with the 
 icon, “Err”, and an error number on the primary display. See Table 7. These 
error conditions disable testing and, if necessary, stop a running test. 
Table 7. Error Codes 
Error Condition 
Code 
Solution 
Self-Test Fails 
Return the tester to a Fluke Service Center. 
Over-Temp 
Wait while the tester cools down. 
Fault Voltage 
Check the installation, in particular, the voltage 
between N and PE. 
Excessive Noise 
Switch off all appliances (Loop, RCD 
measurements) and move the earth stakes 
(earth measurement). 
Excessive Probe 
Resistance 
Put the stakes deeper into the soil. Tamp 
down the soil directly around the stakes. Pour 
water around the stakes but not at the earth 
ground under test. 
 
apx032f.eps 