Fluke 1652C05-TPOLEKITVDE-tester 4426016 User Manual

Product codes
4426016
Page of 66
1652C/1653B/1654B 
Users Manual 
52 
No Trip and Hi Current Modes RCD/FI 
Mains Input Voltage Range 
100 - 500 V ac (50/60 Hz) 
Loop Impedance: phase to earth 
Input Connection (soft key selection)
Line impedance: phase to neutral 
Limit on Consecutive Tests 
Automatic shutdown when internal 
components are too hot. There is also a 
thermal shutdown for RCD tests. 
Maximum Test Current @ 400 V 
20 A sinusoidal for 10 ms 
Maximum Test Current @ 230 V 
12 A sinusoidal for 10 ms 
 
Range Resolution 
Accuracy
[1]
 
10 
Ω
 0.001 
Ω
 
Hi Current m mode: 
±
(2 % + 15 digits) 
No Trip mode: 
±
(3 % + 6 digits) 
20 
Ω
 0.01 
Ω
 
Hi Current mode: 
±
(2 % + 4 digits) 
No Trip mode: 
±
(3 %) 
200 
Ω
 0.1 
Ω
 
Hi Current mode: 
±
(2 %)
 
2000 
Ω
 1 
Ω
 
±
6 %
[2]
 
Notes 
[1]  Valid for resistance of neutral circuit <20 
Ω
 and up to a system phase angle of 30 
°
. Test leads 
must be zeroed before testing. 
[2]  Valid for mains voltage >200 V. 
Prospective Earth Fault Current Test (PSC/I
K
Computation 
Prospective Earth Fault Current (PEFC/I
K
) or Prospective 
Short Circuit Current (PSC/I
K
) determined by dividing 
measured mains voltage by measured loop (L-PE) 
resistance or line (L-N) resistance, respectively. 
Range 
0 to 10 kA or 0 to 50 kA 
(See Power-On Options earlier in this manual) 
Resolution Units 
I
K
 <1000 A 
1 A 
Resolution and Units 
I
K
 >1000 A 
0.1 kA 
Accuracy 
Determined by accuracy of loop resistance and mains 
voltage measurements.