Gossen Metrawatt M817R Mains-analysis device, Mains analyser M817R Data Sheet

Product codes
M817R
Page of 228
5-16
Nominal Values, continued
Frequency 
synchronization
Normal frequency synchronization (sync) of the Energy Platform analog signal 
processing circuits is controlled by analyzing the external AC voltage signal input on 
channel A. If sync is lost, then internal sync is automatically selected. The frequency 
used for internal is entered under Frequency on the Nominal Values screen (see page 5-
14). 
Typically, external sync on Ch D is specified if it is known that signal on A is 
inappropriate for use as a sync source (i.e. the frequency is unstable or DC). A low 
distortion 50, 60 or 400 Hz signal can be input to Ch D and used as a sync signal.
NOTE: Failure to connect an external sync or to select an internal sync during current 
monitoring will result in erroneous data being collected and displayed.
Frequency 
tracking
Tracking determines how the phase locked loop circuitry responds to sudden changes 
in frequency. 
Standard mode: On a normal power quality application [Standard], the system is 
designed to “ride through” voltage changes to keep tracking the fundamental frequency 
even in the absence of voltage. 
Fast mode: When operating on a local generator [Fast], the frequency will vary more as 
loading changes occur. The Fast mode allows the instrument to follow these changes 
more closely. 
Internal mode: In cases where no reference signal is available (such as DC monitoring), 
the Internal setting is used to set the sample rate.