Operating Manual GMH 3530 
page 4 of 14
 
 
Display and control elements 
5.1  Display elements 
 
 
Main display: pH-value, redox-value (mV, mV
H
), rH  
value or user prompt 
 
Display elements to show minimum/maximum/ 
memorized measuring value 
 
Display of measuring units 
 
Warning signal (low battery or recalibration prompt)  
 
ATC-arrow: indicates if temperature sensor has been 
connected, i.e. if automatic temperature 
compensation 
is active, when operating in 
the pH, mV
or rH mode. 
 
ok-arrow: indicates that measuring value has been sta-
ble for a longer period of time 
 
CAL-arrow: indicates that device is being calibrated at 
the moment, when operating in the pH 
mode. 
 
Secondary display: measuring value, temperature 
or user prompt 
5.2  Pushbuttons 
 
 
On/off key 
 
 
min/max when taking measurements: 
press shortly: min. or max. meas. value so far will be 
displayed 
press for 2 sec.: the min. or max. value will be deleted 
Configuration: to enter values, or 
change settings 
 
CAL
: for ’pH’ mode only: 
press shortly: display state of electrode condition and 
calibration data 
press for 2 sec: start pH-calibration 
 
Set/Menu
press (Set) shortly: for ’pH’ and ’mV
H
: manual tempera-
ture input if no temperature probe is con-
nected additionally for ’rH’: manual input of 
pH value.  
press (Menu) for 2 sec: configuration will be activated 
 
Store/Quit
measuring: holds and memorizes current meas. value 
(’HLD’ in display) 
Configuration: enter setting, return to measuring 
5.3  Connections 
 
interface: connection for galv. Isolated interface adapter  
(accessory: GRS 3100, GRS3105 or USB3100)
 
BNC-socket: connection for pH- or redox-electrode 
Mini-DIN-socket: connection for Pt100-temperature probe  
(4-wire connection; 
2-wire also possible, but may result in additional mea-
suring faults due to cable) 
The mains socket is located at the left side of the instrument 