Extech RE300 Waterproof ORP Meter RE300 User Manual

Product codes
RE300
Page of 5
ExStik
TM
  RE300-EU_ENG   V2.2   02/08 
Operation 
Getting Started 
Battery Activation: An insulating tab is located in the battery compartment and prevents 
battery drain during storage. It must be removed before use.  
White KCL crystals may be present in the cap. These crystals will dissolve in the soak or 
they can be simply rinsed with tap water  
When the RE300 is turned on, it enters the Automatic Calibration mode. ‘SELF’ and ‘CAL’ 
will appear on the LCD while calibration is in progress. After the calibration is complete, 
‘SELF’ and ‘CAL’ will extinguish and both the main display and bargraph will read ORP in 
mV units. The bargraph reads 0mV (far left), 500mV (center), and 1000mV (far right). 
 
ORP Testing 
1. 
Remove cap from the electrode and press the ON/OFF button to switch the RE300 
on. 
2. 
Dip the electrode1/2” to 1” into the test solution. Stir once and let the reading stabilize 
for 1 to 3 minutes. 
3. 
Note the ORP or press MODE to freeze the reading The HOLD display icon will 
appear along with the held reading. The held reading will be stored in the next 15-
reading storage location (see below). Press the MODE key to return to normal 
operation. 
4. 
Press the ON/OFF button to turn the RE300 off. If you do not press a button for 10 
minutes the RE300 shuts off automatically. 
Data Hold 
Press the MODE button to freeze the current reading. The HOLD display icon will appear 
along with the held reading. The held reading will be stored in the next 15-reading storage 
location (see below). Press the MODE key to return to normal operation. 
15-Reading Measurement Storage 
1. 
Press the MODE button to store a reading. The LCD will briefly display the memory 
location number and then the value stored (Data Hold will activate).  
2. 
Press MODE again to return to normal 
operation. Repeat step 1 above to store 
the next reading and so on. 
3. 
If an attempt is made to store more than 
15 readings, the stored readings (starting 
with the first reading) will be overwritten. 
Recalling Stored Readings 
Note: First ensure that the HOLD symbol is not displayed. If it is, exit the HOLD function by 
pressing the MODE button. 
1. 
Press the CAL button once and then press the MODE button immediately after CAL is 
displayed; the location number (1 through 15) will flash. If the CAL mode is 
accidentally accessed (display flashing), press the CAL button again to exit. 
2. 
The last stored reading taken will be displayed first. To advance through the stored 
readings, press the MODE button. The location number is displayed first, followed by 
the reading stored in that location. 
3. 
To exit the storage mode, press the CAL button and the ExStik
TM
 will return to normal 
operation.