Greisinger GDH 12 AN Digital Vacuum Barometer for Absolute Pressure 600524 Leaflet

Product codes
600524
Page of 2
H18.0.03.6C-01
page  2 of 2
Safety Requirements:
This device has been designed and tested in accordance with the safety regulations for electronic devices.
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and special safety
advises given in this manual will be adhered to when using the device.
1.  Mains operation:
  When using a power supply device please note that operating voltage has to be 9 to 12 V DC.
Do not apply overvoltage!! Cheap 12V-power supply devices often have excessive no-load voltage. We, therefore, re-
commend using regulated voltage power supply devices. Trouble-free operation is guaranteed by our power supply GNG10.
  Prior to connecting the power supply to the mains make sure that the operating voltage stated at the power supply is
identical to the mains voltage.
2.  If device is to be connected to other devices (e.g. via serial interface) the circuitry has to be designed most carefully. Internal
connection in third party devices (e.g. connection GND and earth) may result in not-permissible  voltages impairing or
destroying the device or another device connected.
Warning: If device is operated with a defective mains power supply (short circuit from mains voltage to output voltage) this
may result in hazardous voltages at the device (e.g. sensor socket at interface)
3.  Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to any other climatic
conditions than those stated under "Specification".
If the device is transported from a cold to a warm environment condensation may cause in a failure of the function. In such a
case make sure the device temperature has adjusted to the ambient temperature before trying a new start-up.
4.  If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be marked
accordingly to avoid re-starting .
Operator safety may be a risk if:
  there is visible damage to the device
  the device is not working as specified
  the device has been stored under unsuitable conditions for a longer time
In case of doubt, please return device to manufacturer for repair or maintenance.
What to observe during operation:
a)  When to replace battery:
Make sure to apply correct operating voltage as wrong voltage will lead to measuring inaccuracies. As soon as "BAT" is
displayed, battery is used up and needs to be replaced.
If the indication is ignored (e.g. device switched on inadvertently for an extended period of time) and battery is very low, the
battery voltage may be too low to trigger BAT display. The values in the display may, however, seem to be correct (although
they no longer correspond to the analog output voltage) !
In case of measuring faults make it a rule to first of all check battery and replace battery if necessary !
b)  Both measuring device and sensor have to be treated carefully and should only be used in accordance with the
specifications (do not throw, bump device etc.). Plug and socket have to be protected from soiling.
c)  The pressure sensor may be disconnected - do not tear at cable when doing so!
When connecting the sensor the arrow on the plug has to point upwards; make sure to enter the plug centrally into the
socket. Plug has to be level when being entered into socket, do not tilt ! In the correct position the plug will easily slide into
the socket.
An inaccurate or tilted position when trying to connect the sensor may lead to the connector pins being spoiled by bending or
being broken. => Plug can no longer be used and will have to be replaced.
Please note: When exchanging the sensor, the device will have to be re-calibrated for the new sensor!
d)  Connection of pressure tube:
For absolute pressure measurement:
  Connect plastic tube with internal diameter of 4 mm to connecting terminal "A";
connection "B" will not be used!
As the sensor connection terminal is made of plastic, they may break when being handled inexpertly (e.g. bending etc.). To
disconnect sensor you should simultaneously tear and turn at the plastic tube.
If tube has to be changed frequently, it is good practice to connect a short tube permanently at the sensor and to use an
adapter (accessories) at the open end.
e)  The device and the sensor will be calibrated before leaving our works.
If the permissible pressure has been exceeded (overload), recalibration of the device and sensor may be required.
In case you are not in a position to carry out recalibration yourself, please return both sensor and device to our works.