A red indicator (Fig. 2) on the left side of 
the instrument indicates the automatic 
change of range. When it is lit, multiply the 
reading by 10 to obtain the result. 
Measurement input terminals: These 
accept safety plugs 4mm in diameter. 
Other indicators:  
A  green  indicator  (Fig.  1)  on  the  right  side 
of  the  instrument  indicates  that  the  speed 
of the magneto is correct. When it is lit, the 
voltage  between  the  terminals  of  the 
instrument  is  250V,  500V,  or  1000VDC 
(no-load),  depending  on  the  setting  of  the 
switch. 
 
3.  USE AND CHARACTERISTICS: 
3.1.  RECOMMENDATIONS 
The  needle  must  indicate  0  on  the  600-scale  when  at  rest;  if  not,  adjust  it 
using the black screw in the centre (see picture above). 
Do  not  connect  the  instrument  to  the  terminals  of  a  circuit  of  which  the 
voltage exceeds 600V
AC
 or 
DC
Before any resistance measurement, check that the circuit is not live (switch 
set to 500V before connection). 
After  each  insulation  resistance  measurement,  let  the  circuit  discharge 
(needle  on  0V;  this  takes  a  few  seconds)  before  disconnecting  the 
instrument.