Chauvin Arnoux CA 6503 Insulation measuring device, P01132504 User Manual

Product codes
P01132504
Page of 57
English 
C.A 6503 
21 
 
Overloads - protection: 
 
M
Ω
 ranges: 600VAC/DC max. 
Compliance with international standards:  
 
Electrical safety in accordance with IEC 61010-1 (Ed. 2 of 2001). 
Electrical safety in accordance with IEC 61010-2-031 (Ed. 2 of 2002). 
Measurement in accordance with IEC 61557 (Ed. 2 of 2007), parts 1 and 

Rated maximum voltage: 600V  
Rated characteristics: measurement category III, 300V, or measurement 
category II, 600V with respect to earth,  
degree of pollution 2. 
 
Electromagnetic compatibility: 
The instrument complies with EMC requirements as per 
NF EN 61326-1 : 2006. 
 
5.  WARRANTY 
Our  warranty  applies,  except  as  otherwise  expressly  stipulated,  for  twelve 
months  counting  from  the  date  of  availability  of  the  equipment  (extract  from 
our General Terms of Sale, communicated on request). 
 
 
6. MAINTENANCE 
6.1.  STORAGE: 
Snap  the  crank  into  the  recess  provided  for  this  purpose;  fit  the  protective 
cover. 
There is nothing that need be removed for prolonged storage.