Honeywell HIH-5031-001 Data Sheet

Page of 8
www.honeywell.com/sensing 
Table 1. Performance Specifications (At 3.3 Vdc supply and 25  C [77  F] unless otherwise noted.) 
Parameter 
Minimum 
Typical 
Maximum 
Unit 
Specific 
Note 
Interchangeability (first order curve) 
   0% RH to 10% RH, 90% RH to 100% RH 
   11% RH to 89% RH 
 
-7 
-3 
 
 
 
 

 
% RH 
% RH 
 
 
Accuracy (best fit straight line) 11% RH to 89% RH 
-3 
 
+3 
% RH 
Hysteresis 
 
 
% RH 
 
Repeatability 
 
0.5 
 
% RH 
 
Settling time 
 
 
70 
ms 
 
Response time (1/e in slow moving air) 
 
 
 
Stability (at 50% RH in 5 years) 
 
1.2 
 
% RH 
Voltage supply 
2.7 
 
5.5 
Vdc 
Current supply 
 
200 
500 
 
Voltage output (1st order curve fit) 
V
OUT
=(V
SUPPLY
)(0.00636(sensor RH) + 0.1515), typical at 25  C 
Temperature compensation 
True RH = (Sensor RH)/(1.0546   0.00216T), T in  C 
Output voltage temp. coefficient at 50% RH, 3.3 V 
 
-2 
 
mV/ C 
 
Operating temperature 
-40[-40] 
See Figure 2. 
85[185] 
C[ F] 
 
Operating humidity (HIH-5030) 
See Figure 2. 
100 
% RH 
Operating humidity (HIH-5031) 
See Figure 2. 
100 
% RH 
 
Storage temperature 
-50[-58] 
 
125[257] 
C [ F] 
 
Storage humidity 
See Figure 3. 
% RH 
Specific Notes: 
1.  Includes stress outside of recommended operating zone. 
2.  Device is tested at 3.3 Vdc and 25  C. 
3.  Non-condensing environment. When liquid water falls on 
the humidity sensor die, output goes to a low rail condition 
indicating no humidity. 
4.  Total accuracy including interchangeability is  3 %RH. 
General Notes: 
  Sensor is ratiometric to supply voltage. 
  Extended exposure to >90 % RH causes a reversible shift 
of 3 % RH. 
  Sensor is light sensitive. For best performance, shield 
sensor from bright light.