Gw Instek GDS-2074A 4-channel oscilloscope, Digital Storage oscilloscope, GDS-2074A Data Sheet

Product codes
GDS-2074A
Page of 220
 
 
GDS-2000A Programming Manual 
106
 
Reference Commands 
 
 
:REF<X>:DISPlay ....................................................... 106
 
:REF<X>:TIMebase:POSition .................................... 106
 
:REF<X>:TIMebase:SCALe ........................................ 107
 
:REF<X>:OFFSet ........................................................ 107
 
:REF<x>:SCALe .......................................................... 108
 
 
 
:REF<X>:DISPlay 
 
Set
Query
 
 
Description 
Sets or queries a reference waveform to be shown 
on the display. A reference waveform must first be 
saved before this command can be used. 
Syntax 
:REF<x>:DISPlay { OFF| ON| ? } 
Parameter 
<X> 
Reference waveform 1, 2, 3 ,4. 
 
OFF 
Turns the selected reference 
waveform off 
 
ON  
Turns the selected reference 
waveform on 
Return parameter 
Returns the status of the selected reference 
waveform. (OFF, ON) 
Example 
:REF1:DISPlay ON
 
Turns on reference1 (REF 1) on the display. 
:REF<X>:TIMebase:POSition  
Set
Query
 
 
Description 
Sets or returns the selected reference waveform 
time base position. 
Syntax 
:REF<X>:TIMebase:POSition { <NRf> | ?} 
Related 
commands 
:REF<X>:DISPlay