REMOTE CONTROL CONFIG 
229 
Configure GPIB  
 
Note  
To use GPIB, the optional module, DS2-GPIB, must be 
installed. Please see page 30 for installation details. 
Connection 
1.  Connect a GPIB cable from a PC to the installed 
GPIB module. 
Configure GPIB 
2.  Press the Utility key. 
Utility
 
 
3.  Press I/O from the bottom menu. 
 
 
4.  Use the Variable knob to set the 
GPIB Address from the side menu. 
This option will only be available 
when the GPIB module is installed. 
 
 
 
Range 
1 ~ 30 
GPIB Constraints 
 
Maximum 15 devices altogether, 20m cable 
length, 2m between each device 
 
Unique address assigned to each device 
 
At least 2/3 of the devices turned On 
 
No loop or parallel connection