Gw Instek GDS-2074A 4-channel oscilloscope, Digital Storage oscilloscope, GDS-2074A Data Sheet

Product codes
GDS-2074A
Page of 261
 
GETTING STARTED 
33 
First Time Use 
 
Background 
This section describes how to connect a signal, 
adjust the scale, and compensate the probe. Before 
operating the GDS-2000A in a new environment, 
run these steps to make sure the instrument 
performs at its full potential.  
1. Power On 
Follow the procedures on the previous page. 
2. Set the Date 
and Time 
Set the date and time. 
3. Reset System 
Reset the system by recalling the 
factory settings. Press the Default key 
on the front panel. For details, see 
page 202. 
Default
 
4. Install Optional 
modules 
There are a number of optional 
hardware modules that can be 
installed, such as the optional 
function generator. 
5. Install Optional 
Software 
Optional software packages may also 
need to be installed. 
6. Connect Probe 
Connect the probe to the Channel 1 input and to 
the CAL signal output (Demo 3 output). This 
output provides a 2Vp-p, 1kHz square wave for 
signal compensation by default. 
Set the probe attenuation to x10 if the probe has 
adjustable attenuation.