GDS-2000A Series User Manual 
116
 
Rerun Segmented 
Acquisition 
1.  To rerun the segments, press the Segments Stop 
key to toggle the mode back to the Segments 
Run
 mode. 
 
 
 
Alternatively, press the Run/Stop 
key again. 
Run/Stop
 
 
2.  Repeat steps 3 and 4 in the section above when 
the segmented acquisition has completed.  
 
The scope is now ready to navigate or analyze the 
acquired segments. 
Navigate Segmented Memory 
 
Background 
After the segmented memory acquisitions have 
been captured you can navigate through each 
segment one at a time. 
Operation 
1.  Press Select Segments from the 
bottom menu. This key will be 
available in the Stop mode. 
 
 
2.  To navigate to the segment of 
interest, press Current Seg from the 
side menu and use the Variable 
knob to scroll to the segment of 
interest. 
 
Alternatively, the Set to Minimum 
and Set to Maximum keys can be 
used to jump to the first and last 
segment respectively.