The waveform does not update (frozen). 
 
Press the Run/Stop key to unfreeze the waveform. See page 68 for 
details. 
If this does not help, the trigger mode might be set to Single. Press 
the Single key to exit Single mode. See page 68 for Single trigger 
details. 
The probe waveform is distorted. 
 
You might need to compensate the probe. For details, see page 244.  
Autoset does not catch the signal well. 
 
The Autoset function cannot catch signals under 30mV or 20Hz. 
Please use the manual operation. See page 66 for Autoset details. 
I can’t save files to the internal memory. 
 
If a USB stick is inserted into one of the USB slots and you wish to 
save to the scope internal memory, press the Utilities key and set 
the file path to internal memory.  
The display image printout is too dark on the background. 
 
Use the Ink Saver function which reverses the background color. 
For details, see page 219. 
The date and time settings are not correct. 
 
For date and time setting details, please see page 170. If it does not 
help, the internal battery controlling the clock might be worn out. 
Contact your dealer or GWInstek.