Voltcraft UTS-1980 Function Generator, Frequency Generator UTS-1980 Data Sheet

Product codes
UTS-1980
Page of 33
 
 
6
Note:   Time base Error: 
10ppm(at normal temperature)   
Trigger Error
0.3% when SNR(Signal-to-Noise Ratio)>40dB   
1.4 Dynamic Range:         CHA   50Vrms~1.5Vrms  
                         CHB   -13dBm~+13dBm  (100MHz-1GHz) 
                                -7dBm~+13dBm    (1GHz-2.7GHz)  
                             
1.5 Max. undamage voltage:  CHA   35Vpp   
                                                  CHB      3Vpp 
1.5 Input impedance:        CHA   1MΩ//40pF 
                                                  CHB      50Ω 
1.6 Time base frequency:    10MHz 
1.7 Resolution:           ±(100ns/Gate times)x measured signal frequency                                 
2. Basic Operation 
2.1 Please check voltage and fuse before the power is on. Turn on main power switch
,
36
and then turn on FC power switch(
,
39
which is on the rear panel of 
instrument and then preheat 20 minutes.   
2.2 After the power is on, the display window(
,
1
  of FC indicates “0000000”, the 
indicator of CHA will be on. The strobe time is 1s and the indicator of strobe is 
on. The indicators of Hold and Low pass filter (LPF-A) will be off. 
2.3 LPF-A key(
,
6
: bandwidth about 100kHz at 3dB, Open LPF (indicator of 
LPF-A will be on) to eliminate the HF disturb of the LF signal below 100 kHz 
which will make the measured result more reliable. 
2.4 Channel key(
,
5
: It is used to select the measured channel. CHA can be used to 
measure 1Hz ~ 100MHz signal frequency. CHB can be used to measure 
100MHz~2.7GHz signal frequency. 
2.5 For CHA,   
a. When frequency is <100 kHz, please turn on the low pass filter (LPF-A).   
b. When the amplitude of frequency is >5V, please use external attenuators, such 
as oscilloscope-probes. 
2.6 Strobe time select key(
,
4
: It should choose a suitable strobe time if a suitable 
resolution is needed. Once the period of the measured signal is more than strobe 
time, the measurement time is regarded as a period time of the measured signal. 
2.7 Data Hold(
,
3
: Press it down, the Hold indicator will be on. Then the FC will 
stop measurement. The display window will always keep the last measured data 
even the signal frequency is changed or test dot is broken. Press Hold again, its 
indicator will be off. Then FC re-starts to measure and the display window will 
record new frequency measurement data.   
2.8 Reset(
,
2
: It is used to reset the FC.