Metrel MD 9020 Digital-Multimeter, DMM, 20991442 Data Sheet

Product codes
20991442
Page of 40
MD 9020 Electrical & Electronic DMM Maintenance 
 
16 
6 Specification 
 
GENERAL SPECIFICATIONS 
 
Display:  
 
 
3-3/4 digits 4000 counts LCD display 
 
Polarity:  
 
 
Automatic 
  
Update Rate:    
 
3 per second nominal 
 
Operating Temperature:  
0
C to 40C 
 
Relative Humidity :  
Maximum relative humidity 80% for temperature up to 31
C decreasing 
linearly to 50% relative humidity at 40
C 
 
Altitude:  
 
 
Operating below 2000m 
 
Pollution degree:  
 
 
Storage Temperature:  
-20
O
C to 60
O
C, < 80% R.H. (with battery removed) 
 
Temperature Coefficient:  
nominal 0.15 x (specified accuracy)/
O
C @(0
O
C -18
O
C or 28
O
C -40
O
C), or 
otherwise specified 
 
Sensing:  
 
 
Average sensing for MD 9020. 
  
Safety:  
The meter (all versions) is protected, against the users, by double 
insulation per UL3111-1(1994), CSA C22.2 No. 1010-1-92, EN61010-
1(1995) and IEC61010-1(1995) to CAT II 1000V & CAT III 600V
 
Standard CE version series Terminals (to COM) ratings:
 
  V : 
 
Category II 1000 Volts AC & DC, and Category III 600 Volts AC & DC 
  A / mAA : 
Category II 250 Volts ac and 150 Volts dc 
 
Protection-reinforced CE + UL version series Terminals (to COM) ratings: 
  V : 
 
Category II 1000 Volts AC & DC, and Category III 600 Volts AC & DC 
  A / mAA : 
Category III 500 Volts AC and 300 Volts DC 
 
Overload Protections 
 
Standard CE version: 
  A & mA :  
0.5A/250V F Fuse;  
  A: 
 
15A/250V F Fuse;  
  V: 
 
1050Vrms, 1450Vpeak;  
  , & Others 
: 600VDC/VAC rms 
   
Protection-reinforced CE + UL version: 
  A & mA :  
0.63A/500V, IR200kA, F Fuse 
  A :  
 
12.5A/500V, IR20kA, F Fuse 
  V :  
 
1050Vrms, 1450Vpeak;  
  , & Others:   600VDC/VAC rms 
 
Transient protection:    
6.5kV (1.2/50
s surge) 
 
Power Supply:  
 
1.5V AAA Size (NEDA 24A or IEC LR03) battery X 2 
 
Power Consumption:   
3.2 mA typical