Rigol DM3058E Digital-Multimeter, DMM, DM3058E Data Sheet

Product codes
DM3058E
Page of 8
 
                 2 
                                                           
Specifications                                                     
 
 
DC Characteristics   
Accuracy  ± (%  of  reading  + % of  range)
[1]
 
Function 
Range
[2]
 
Test current or 
Load voltage 
1 Year 
23± 5 
Temperature 
Coefficient 
0 - 18 
28
 - 50 
DC Voltage 
200.000 mV 
 
0.015 + 0.004 
0.0015 + 0.0005 
2.00000 V 
 
0.015 + 0.003 
0.0010 + 0.0005 
20.0000 V 
 
0.015 + 0.004 
0.0020 + 0.0005 
200.000 V 
 
0.015 + 0.003 
0.0015 + 0.0005 
1000.00 V
[4]
 
 
0.015 + 0.003 
0.0015 + 0.0005 
DC Current 
200.000 μA 
<8 mV 
0.055 + 0.005 
0.003 + 0.001 
2.00000 mA 
<80 mV 
0.055 + 0.005 
0.002 + 0.001 
20.0000 mA 
<0.05 V 
0.095 + 0.020 
0.008 + 0.001 
200.000 mA 
<0.5 V 
0.070 + 0.008 
0.005 + 0.001 
2.00000 A 
<0.1 V 
0.170 + 0.020 
0.013 + 0.001 
10.0000 A
[5]
 
<0.3 V 
0.250 + 0.010 
0.008 + 0.001 
Resistance
[3]
 
200.000 Ω 
1 mA 
0.030 + 0.005 
0.0030 + 0.0006 
2.00000 kΩ 
1 mA 
0.020 + 0.003 
0.0030 + 0.0005 
20.0000 kΩ 
100 μA 
0.020 + 0.003 
0.0030 + 0.0005 
200.000 kΩ 
10 μA 
0.020 + 0.003 
0.0030 + 0.0005 
2.00000 MΩ 
1 μA 
0.040 + 0.004 
0.0040 + 0.0005 
10.0000 MΩ 
200 nA 
0.250 + 0.003 
0.0100 + 0.0005 
100.000 MΩ 
200 nA || 10 MΩ 
1.75  + 0.004 
0.2000 + 0.0005 
Diode Test 
2.0000 V
[6]
 
1 mA 
0.05 + 0.01 
0.0050 + 0.0005 
Continuity 
Test 
2000 Ω 
1 mA 
0.05 + 0.01 
0.0050 + 0.0005 
Remarks 
[1]  Specifications are for 0.5 hour warm-up, “Slow” measurement rate and calibration temperature 18℃ ~ 28℃.   
[2]  20% over range on all ranges except for DCV 1000 V, ACV 750 V, DCI 10 A and ACI 10 A.   
[3]  Specifications are for 4-wire measure or 2-wire measure under “REF” operation. ± 0.2 Ω of extra errors will be 
generated if perform 2-wire measure without “REF” operation. 
[4]  Plus 0.02 mV of error per 1 V after the first ±500 VDC. 
[5]  30 seconds OFF after 30 seconds ON is recommend for the continuous current that higher than DC 7 A or AC RMS 
7 A.   
[6]  Accuracy  specifications are only for voltage  measuring  at input terminal. The typical value of current under 
measure is 1 mA. Voltage drop at the diode junction may vary with current supply.        
 
DC Voltage 
Input Resistance 
200 mV and 2 V    10 MΩ or >10 GΩ selectable 
(Input signals that exceed ± 2.5 V in these ranges will pass the 100 kΩ (typical) clamp 
resistance.) 
20 V, 200 V and 1000 V        10 MΩ ± 2% 
Input Bias Current 
<90 pA, 25℃ 
Input Protection 
1000 V on all ranges   
CMRR  (common mode 
rejection ratio) 
120 dB (For the 1 kΩ unbalanced resistance in LO lead, maximum ± 500 VDC) 
NMRR  (normal  mode 
rejection ratio) 
60 dB at “slow” measurement rate 
20  dB  are added  if  open the “AC filter”  (Settling time  will add  0.35  s  while  source 
impedance nears zero) 
Resistance 
Testing Method 
4-wire resistance or 2-wire resistance optional 
For current source refer to LO input 
Open-circuit Voltage 
P/F in <8 V 
Maximum Lead 
Resistance 
(4-wire resistance) 
10% of ranges, on the range of 200 Ω, 1 kΩ, for each lead 
1 kΩ, on all other ranges, for each lead 
 
Input Protection 
1000 V, on all ranges