OFF 
 
The menu item 5:Math is off.
Lo Limit   
Lower limit.
 
 
If the measured value is smaller than the Lo  
 
  Limit value an acoustic warning sounds and 
“Lo  
 
limit is displayed.
Hi Limit   
Higher limit.
 
 
If the measured value is greater than Hi Limit  
 
 
value an acoustic warning occurs and „Hi limit“  
 
 
is displayed.
Offset 
 
An offset value can be set for all measurement  
 
 functions 
6
 to 
15
 
a)  Apply the offset value to the terminals.
 
b)  Choose menu item 5:Math.
 
c)  Select submenu „Offset“ by pressing   
23
.
 
d)  Open the submenu with MENU 
18
, the    
 
 
offset value applied will be displayed. 
 
e)  Accept the offset by pressing MENU 
18
.
 
f)  Return to measurement function, the display  
 
 
shows 0,00…., , the dimension and „Os“.
 
g)  Now you can connect the value to be  
 
 
 
measured. It is compared the calibrated value  
 
 
and the deviation is displayed.
In order to delete the stored offset:
 
a)  Choose menu item 5:Math.
 
b)  Select submenu „Off“ by pressing   
23
.
 
c)  Accept by pressing MENU 
18
, return to    
 
 
measurement function, the measured value is  
 
 
displayed without offset.
or  
d)  select another measurement function. 
 
 
The offset value will not be stored when the  
 
 
measurement function is changed.
6:logger
Analysis of different characteristics of the measured values
Start 
 
The test series is started. According to the  
 
 
selected sampling rate in „0:Time“ every xx  
 
 
second a reading is taken and stored.
Stop 
 
The test series is stopped.
Dump 
 
The test series is shown on the display. Each time
  
 
button ENTER 
22
 is pressed the next value of the
 
   stored test series is displayed.
7:com
In this menu the baud rate can be chosen. Either 9600 baud or 
19200 baud are available. The remaining interface parameters 
cannot be changed.
interface parameters (adjustable)
Rs Off   
The interface is switched off 
Rs19200  
19,200 baud
Rs9600   
9,600 baud
interface parameters (not selectable)
 
no parity bit
 
8 data bits
 
1 stop bit
Xon-Xoff Xon-Xoff
  every transmission of a character takes 1 ms. se-
lecting a sampling rate of 0.01 sec requires a baud 
rate of 19 200.
8:cal
This menu is saved by password. In order to guarantee exact 
measurements the HM8112-3 is calibrated. Calibration may only 
be done with adequate precision reference sources. For this 
purpose the password can be orderd at HAMEG GmbH (Phone.: 
(+49) 06182-800-500 or via E-Mail: service@hameg.com).
          s c a n n e r  c a r d  H o 1 1 2
Attention: 
After receiving the password any warranty claims of HAmeg 
GmbH concerning the compliance with the technical specifi-
cations of the instrument become void.
9:mux
For the future implementation of a scanner/test point switch.
10.8  measurement inputs
For connection measurement signals the HM8112-3 features 4 
safety connectors on the front panel. Depending on the measu-
rement function chosen, the active terminals will be illuminated.
  the terminals on the front panel are safety con-
nec-tors and the regulations have to be observed.
if connecting dangerous voltages to the input termi-
nals 
24
 and 
26
 all relevant safety regulations are to 
be observed.
DC voltage must be floating!
AC voltage must be floating by use of a safety isola-
ting transformer.
Attention!
Voltages exceeding one of the following values are 
regarded potentially dangerous or even lethal:
1
st
 
30 V
rms
2
nd
 
42.4 V
peak
3
rd
 
60 V
Dc
connecting higher voltages is only allowed by skilled 
personnel who are familiar with the dangers incur-
red. the relevant safety regulations are to be strictly 
observed!
24
  V/sense (4mm safety sockets)
Connection of measuring cables for:
–  voltage measurement 
–  frequency measurement
–  4 wire resistance measurement (SENSE)
–  continuity test
–  temperature measurement by a thermocouple
–  4 wire temperature measurement by a P-temperature 
sensor (SENSE)
the maximum voltage between Hi and lo case 
(ground) must not exceed 850 V-
peak
 or 600 V
Dc
.
the maximum voltage between lo and case 
(ground) may not exceed 250 V
rms
!
c o n t r o l  e l e m e n t s  a n d  d i s p l a y          
24
25
26