Users Manual 
2-4 
Selecting the Line Voltage 
The Meter operates on four different input line voltages. The selected line-voltage setting 
is visible through the window in the line-fuse holder on the Meter’s rear panel.  
1.  Unplug the power cord. 
2.  Insert a small screwdriver blade into the narrow recess to the left of the fuse 
holder and pry it to the right until the holder pops out. See Figure 2-1. 
3.  Remove the voltage-selector block from the fuse holder. 
4.  Rotate the selector block until the desired voltage rating faces outward. 
5.  Replace the selector block back into the fuse holder. 
6.  Install the fuse holder back into the Meter and reconnect the power cord. 
Changing the line-voltage setting may require a different line-power fuse for proper 
operation. 
Replacing the Fuses 
The Meter uses one fuse to protect the line-power input and two fuses to protect the 
current-measurement inputs. 
Line-Power Fuse 
The Meter has a line-power fuse in series with the power supply. Table 2-1 indicates the 
proper fuse for each of the four line-voltage selections. The line-power fuse is accessed 
through the rear panel. 
1.  Unplug the power cord. 
2.  Insert a small screwdriver blade into the narrow recess to the left of the fuse 
holder and pry it to the right until the holder pops out. See Figure 2-1. 
3.  Remove the fuse and replace it with a fuse of an appropriate rating for the 
selected line-power voltage. See Table 2-1. 
4.  Replace the selector block back into the fuse holder. 
XW
 Warning 
To avoid electric shock or fire, do not use makeshift fuses or 
short-circuit the fuse holder. 
Table 2-1. Line Voltage to Fuse Rating 
Line Voltage Selection 
Fuse Rating 
100 / 120 
0.125 A, 250 V (slow blow) 
220 / 240 
0.063 A, 250 V (slow blow)