Bias Setting 
 
Background 
Set internal or external bias voltage. 
Panel Operation 
1.  Use the arrow menu keys (F1/F2) 
to move the cursor to INT.B or 
EXT.B. 
 
 
 
 
2.  Press INT (F3) to use internal 
biasing. 
 
 
3.  Press EXT (F4) to use external 
biasing. 
 
 
4.  Use the arrow menu keys to 
highlight OFF/ON.  
 
 
 
 
5.  Press ON (F3) to turn bias voltage 
on. 
 
 
6.  Press OFF (F4) to turn bias voltage 
off. 
 
 
   LCR-800 
User 
Manual 
58
 
Range Setting 
 
Background 
The range can be selected from 1 to 4. Different 
ranges should be used for different components 
and component values and to ensure accurate 
readings. 
 Component 
 Inductor 
Capacitor 
Resistor 
Range1 
1~16mH/f 1.6~25uF/f 6.25~100Ω 
Range2 
16~256mH/f 100~1600nF/f  0.1~1.6kΩ 
Range3 
256~4100mH/f 6.4~100nF/f  1.6~25.6kΩ 
Range4
*
 
4.1~65H/f 400~6400pF/f 25.6~410kΩ 
 
f = test frequency in kHz 
 
*
  This range is not used above 20 kHz  
Panel Operation 
1.  Use the arrow menu keys (F1/F2) 
to move the cursor to RANGE. 
 
 
 
 
2.  Press F3 (UP) to increase the range 
or F4 (DOWN) to decrease the 
range. 
 
Constant Voltage Setting 
 
Background 
Constant voltage is usually used when a set 
voltage is needed. For details about constant 
voltage, see page 39. 
Panel Operation 
1.  Use the arrow menu keys (F1/F2) 
to move the cursor to C.V.