Gw Instek LCR-816 Digital-Multimeter, DMM, LCR-816 Data Sheet

Product codes
LCR-816
Page of 91
  
PROGRAMMING 
103 
AVERAGE 
Command/Query 
 
Sets the average number from 1~255. The average number indicates 
how many test samples are used to create an averaged test result. 
Syntax STEP:AVER 
<variable><^END^M>or<^J^M> 
Parameter 
<variable>  
Average 
number 
1.00~255. (integer 
values) 
1~255 
Note 
Ensure the number has a total of 4 characters. If a 
number does not use 4 characters, use a “.” and 
additional zero’s (0) to “pad out” the number. 
Example 10 = 10.0 
Example STEP:AVER 
255.<^END^M> 
 
Average is set to 255 samples. 
Query Syntax 
STEP:AVER?<^END^M>or<^J^M> 
Return String 
<string> 
Current average setting 
STEP:AVER <variable><^END>  
(<variable>= 1.00 ~255.)  
Returns the average 
number. 
Query Example 
STEP:AVER?<^END^M> 
STEP:AVER 255.<^END> 
 
The average number is currently 255. 
RECALL CALIBRATION 
Command 
 
Recalls the calibration settings from memory. A return string 
indicates if the command was successful. 
Syntax STEP:RECA<^END^M>or<^J^M> 
Return String 
<string> Recall 
calibration 
RECA:OK<^END>  
Successful 
   LCR-800 
User 
Manual 
104
 
Example 
STEP:RECA<^END^M>or<^J^M> 
RECA:OK<^END> 
 
Calibration was recalled successfully 
BAUD RATE 
Command 
 
Sets the baud rate of the RS232 connection. 
Syntax COMU:<value><^END^M>or<^J^M> 
Parameter 
<value>  
Baud 
rate 
9600  
9600 
19.2  
19200 
38.4  
38400 
57.6  
57600 
1152  
115200 
Return String 
<string> Baud 
rate 
COMU:<value><^END> 
<value>= baud rate  
 
Returns the baud rate 
setting. 
Query Example 
COMU:1152<^END^M> 
COMU:1152<^END> 
 
The baud rate is set to 115200. 
MODEL NUMBER 
Query 
 
This query returns the model number of the LCR-800.  
Query Syntax 
COMU:MONO<^END^M>or<^J^M> 
Return String 
<string> Model 
number 
COMU:MONO:816.<^END>  
LCR-816 
COMU:MONO:817.<^END> 
LCR-817