Gw Instek LCR-816 Digital-Multimeter, DMM, LCR-816 Data Sheet

Product codes
LCR-816
Page of 91
  
INTERACE 
119 
Handler Timing 
 
Background 
The handler timing characteristics are described in 
the timing diagram. Times T1 to T6 are described 
in the relevant tables. 
 
 Timing 
Characteristics 
Trigger Pulse Width 
T1 
MIN 
MAX 
 
 5us 
~  
Measurement start delay 
time 
T2 MIN 
MAX 
 
 140us 
~  
/O_EOM Delay Time After 
Data Output 
T3 MIN 
MAX 
 
 5us 
~  
Calculation and binning time  T4 
  MIN 
MAX 
 
 6ms 
~  
   LCR-800 
User 
Manual 
120
 
 
Analog 
Measurement 
time 
T5   Slow 
Medium 
Fast 
 0.012kHz 
817ms 
817ms 
817ms 
 0.1kHz 
901ms 
125ms 
125ms 
 
 0.12kHz 
901ms 
105ms 
103ms 
 1kHz 
903ms 
59ms 
27ms 
 10kHz 
873ms 
53ms 
17ms 
 100kHz 
873ms 
53ms 
17ms 
Trigger Wait Time 
After /O_EOM 
Output 
T6   Slow 
Medium 
Fast 
 OFF 
2ms 
2ms 
2ms 
 BIN 
4ms 
4ms 
4ms 
 VALUE 
16ms 
16ms 
16ms 
Binning Accuracy   
 
 Fast 
Medium 
Slow 
LCR_827 0.5% 
0.2% 
0.1% 
LCR_829 0.5% 
0.2% 
0.1% 
LCR_826 0.5% 
0.2% 
0.1%