COMMAND OVERVIEW 
119 
CONFigure:RANGe? 
Returns the current range on the first display.  
Return Parameter: 
DCV: 0 .5(500mV), 5(5V), 50(50V), 500(500V), 1000(1000V) 
ACV: 0.5(500mV), 5(5V), 50(50V), 500(500V), 750(750V) 
ACI: 0.0005(500μA), 0.005 (5mA), 0.05(50mA), 0.5(500mA), 5(5A), 
10(10A)   
DCI: 0.0005(500μA), 0.005 (5mA), 0.05(50mA), 0.5(500mA), 5(5A), 
10(10A) 
RES: 50E+1(500Ω ) 50E+2(5kΩ ), 50E+3(50kΩ ), 50E+4 (500kΩ ), 
50E+5(5MΩ ), 50E+6(50MΩ ) 
CAP: 5E-9(5nF), 5E-8(50nF), 5E-7(500nF), 5E-6(5μF), 5E-5(50μF) 
CONFigure:AUTO 
Sets Auto-Range on or off on the first display. 
Parameter: ON | OFF 
Example: CONF:AUTO ON 
CONFigure:AUTO? 
Returns the Auto-Range status of the function on the 1
st
 display. 
Return Parameter: 0|1, 1=Auto range, 0=Manual range 
Secondary Display: CONFigure2 Commands  
 
CONFigure2:VOLTage:DC 
Sets measurement to DC Voltage on the second display and 
specifies the range. 
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)] 
Example: CONF2:VOLT:DC 5  
Sets the voltage range to 5 volts.