*OPC 
Sets operation complete bit (bit0) in SERS (Standard Event Status 
Register) when all pending operations are completed. 
*PSC? 
Returns power On clear status. 
Return parameter: <Boolean>(0|1) 0= don’t clear, 1=clear 
*PSC 
Clears power On status. 
Parameter: <Boolean>(0|1) 0=don’t clear, 1= clear 
*RST 
Recalls default panel setup. 
*SRE? 
Returns the SRER (Service Request Enable Register) contents. 
*SRE 
Sets SRER contents. 
Parameter: <NR1>(0~255) 
Example: *SRE 7  
Sets the SRER to 00000111. 
*STB? 
Returns the SBR (Status Byte Register) contents. 
Example:*STB? 
        >81 
Returns the contents of the SBR as 01010001. 
*TRG 