Gw Instek GDM-8341 Digital-Multimeter, DMM, GDM-8341 Data Sheet

Product codes
GDM-8341
Page of 158
 
 
GDM-834X Series User Manual 
38
 
Crest Factor Table 
 
Background 
Crest factor is the ratio of the peak signal 
amplitude to the RMS value of the signal. It 
determines the accuracy of AC measurement. 
If the crest factor is less than 3.0, voltage 
measurement will not result in error due to 
dynamic range limitations at full scale. 
If the crest factor is more than 3.0, it usually 
indicates an abnormal waveform as seen from 
the below table. 
Crest Factor Table 
 Waveform 
Shape 
Crest factor 
Square wave 
 
1.0 
Sine wave 
 
1.414 
Triangle sawtooth 
 
1.732 
Mixed frequencies 
 
1.414 ~ 2.0 
SCR output  
100% ~ 10% 
 
1.414 ~ 3.0 
White noise 
 
3.0 ~ 4.0 
AC Coupled pulse 
train 
 
>3.0 
Spike 
 
>9.0