Frequency/Period Voltage Range Settings 
The input voltage range for frequency/period measurements can 
be set to Auto range or to manual. By default, both the period and 
frequency voltage range are set to Auto. 
 
Range 
 Voltage 
500mV, 5V, 50V, 500V, 750V 
Manual Range 
1.  When in the PERIOD or FREQ measurement 
mode, press the 2ND key twice. This will allow 
the secondary display to show the voltage 
range. 
2.  Set the range with the Up and Down keys. The 
AUTO indicator will turn off when a new range 
is selected. 
3.  Press the 2ND key twice more to switch the 
secondary display back to the previous view.  
Autorange 
1.  Press the Auto/Enter key. 
2.  AUTO will be displayed on the screen again.  
Display 
Hz
k
AUTO
Autorange 
indicator
Voltage range 
setting
 
Note 
The 2nd key is only used to toggle the view of the 
second display between the menu function (FREQ 
or PERIOD) and the voltage range. The voltage 
range can actually be set without switching to the 
2ND display.