GDM-834X Series User Manual 
6
 
Safety Guidelines 
 
General 
Guideline 
 CAUTION 
 
Make sure that the voltage input level does not 
exceed DC1000V/AC750V. 
 
Make sure the current input level does not 
exceed 12A. 
 
Do not place any heavy object on the 
instrument. 
 
Avoid severe impact or rough handling that can 
lead to damaging the instrument. 
 
Do not discharge static electricity to the 
instrument. 
 
Use only mating connectors, not bare wires, for 
the terminals. 
 
Do not block or obstruct the cooling fan vent 
opening. 
 
Do not perform measurement at the source of a 
low-voltage installation or at building 
installations (Note below). 
 
Do not disassemble the instrument unless you 
are qualified as service personnel. 
 
Make sure that the COM terminal to earth is 
limited to 500Vpk. 
 
(Note) EN 61010-1:2010 specifies the measurement categories and 
their requirements as follows. The GDM-834X falls under category II 
600V. 
 
Measurement category IV is for measurement performed at the 
source of low-voltage installation. 
 
Measurement category III is for measurement performed in the 
building installation. 
 
Measurement category II is for measurement performed on the 
circuits directly connected to the low voltage installation.