Gw Instek GDM-8341 Digital-Multimeter, DMM, GDM-8341 Data Sheet

Product codes
GDM-8341
Page of 158
 
 
GDM-834X Series User Manual 
70
 
Compare Measurement 
Compare measurement checks to see if the measurement data stays 
between a specified upper (high) and lower (low) limit.
 
Applicable 
measurements 
The compare function can be used with the 
following basic measurement functions: 
ACV, DCV, ACI, DCI, Ω, Hz/P, TEMP,   
Procedure 
1.  Press SHIFT → HOLD.  
2.  The high limit setting appears.  
 
Use the Left and Right arrow keys to navigate 
to the digit to be edited, or to select the decimal 
point.  
 
Use the Up and Down arrow keys to edit the 
selected digit, or to place the position of the 
decimal point.   
 
V
 
3.  Press the Enter key to save the high limit setting 
and automatically go on to the low limit setting.  
4.  Enter the low limit setting in the same fashion 
as the high setting. 
5.  Press the Enter key to confirm the low limit 
settings.  
6.  The compare measurement results will appear 
immediately:   
If the current measurement reading is between