COMMAND OVERVIEW 
129 
[SENSe:]UNIT 
Sets the temperature unit. 
Parameter: C|F 
Example: SENS:UNIT C 
Sets the temperature unit to ˚C. 
[SENSe:]UNIT? 
Returns the temperature unit. 
[SENSe:]FUNCtion[1/2] 
Sets the function for the first or second display. 
Parameter: 
(display1):"VOLT[:DC]", "VOLT:AC", “VOLT:DCAC”, 
"CURR[:DC]", "CURR:AC", “CURR:DCAC”, "RES", "FREQ", 
"PER", "TEMP:TCO", "DIOD", "CONT", “CAP” 
(display2): "VOLT[:DC]", "VOLT:AC", "CURR[:DC]", 
"CURR:AC", "RES", "FREQ", "PER", "NON" 
Example: SENS:FUNC1 "VOLT:DC" 
Sets the 1
st
 display to the DCV function. 
[SENSe:]FUNCtion[1/2]? 
Returns the function displayed on the first or second display. 
Return parameter: 
(display 1): VOLT, VOLT:AC,VOLT:DCAC, CURR, 
CURR:AC,CURR:DCAC, RES, FREQ, PER, TEMP:TCO, DIOD, 
CONT, CAP 
(display 2): VOLT, VOLT:AC, CURR, CURR:AC, RES, FREQ, 
PER, NON