The DMM performance doesn’t match the specifications. 
 
Make sure the device is powered On for at least 30 minutes, within 
18~28˚C. This is necessary to stabilize the unit to match the 
specifications. 
The measured voltage does not match the expected value.  
 
There are a number of reasons why the measured value may not 
match the expected values.  
1. Ensure that all connections are connected securely and have a 
good contact at all times. Poor contacts could result in erroneous 
measurements.  
2. Ensure that the appropriate input resistance has been set in the 
System menu. For 500mv and 5V ranges, the input resistance can be 
set to either 10MΩ or 10GΩ.  
3. When measuring AC voltage or current, the RMS of the voltage 
peak is measured, not the voltage peak. See page 37 for details. 
4. The measurement rate settings can have an effect on the accuracy 
of the measurement. Slow measurements are more accurate, while 
the fast rate is not as accurate. 
5. Ensure that an appropriate range setting is used. If a too-large 
range is used, the resolution or the measurement may be affected. 
For more information, contact your local dealer or GWInstek at 
www.gwinstek.com / marketing@goodwill.com.tw.