Gw Instek GDM-8342GPIB Digital-Multimeter, DMM, GDM-8342GPIB Data Sheet

Product codes
GDM-8342GPIB
Page of 158
 
 
APPENDIX 
149 
Specifications 
The specifications apply when the DMM is warmed up for at least 
30 minutes and operates in slow rate.  
Below are the basic conditions required to operate the DMM within 
specifications: 
 
Calibration: Yearly 
 
Operating Temperature Specification: 18~28˚C (64.4~82.4˚F) 
 
Relative Humidity: 80% (Non condensing) 
 
Accuracy: ± (% of Reading + Digits) 
 
AC measurements are based on a 50% duty cycle. 
 
The power supply cable must be grounded to ensure accuracy. 
 
All specifications are applicable to the main (1
st
) display only. 
General Specifications 
 
Specification Conditions: 
Temperature: 23˚C±5˚C 
Humidity: <80%RH, 75%RH for resistance measurement readings greater than 
10MΩ. 
Operating Environment: (0~50˚C) 
Temperature Range: 0~35˚C, Relative Humidity: <80%RH;  
                                  >35˚C, Relative Humidity: <70%RH 
Indoor use only 
Altitude: 2000 meters 
Pollution degree 2 
Storage Conditions (-10~70˚C) 
Temperature Range: 0~35˚C, Relative Humidity: <90%RH;  
                                  >35˚C, Relative Humidity: <80%RH 
General: 
Power Consumption: Max 15VA 
Dimensions: 265 mm (W) X 107 mm (H) X 302 mm (D) 
Weight: Approximately 2.9 kg