Gw Instek GDM-8342GPIB Digital-Multimeter, DMM, GDM-8342GPIB Data Sheet

Product codes
GDM-8342GPIB
Page of 158
 
 
OPERATION 
35 
Select the Voltage Range 
The voltage range can be set automatically or manually.  
 
Auto Range 
To turn the automatic range selection On/Off, 
press the AUTO key.  
Manual Range 
Press the Up or the Down key to select the 
range. The AUTO indicator turns Off 
automatically. If the appropriate range is 
unknown, select the highest range. 
Selectable Voltage 
Ranges 
 Range 
Resolution 
Full scale 
500mV 
10μV 
510.00mV 
5V 
0.1mV 
5.1000V 
50V 
1mV 
51.000V 
500V 
10mV 
510.00V 
750V (AC) 
100mV 
765.0V 
1000V (DC) 
100mV 
1020.0V 
 Note 
For further details, please see the specifications on 
page 149. 
 Note 
DC voltages with AC components cannot be 
accurately measured if the DC+AC component 
exceeds the ADC dynamic range for the selected 
DC range. Any voltage exceeding the ADC dynamic 
range will be clipped at the upper/lower range 
limit. Under these conditions the range that is 
chosen with the Auto range function may be too 
small.