GDM-834X Series User Manual 
86
 
example, if 102000 counts are logged, 3 files will 
be created: GW000\GW000-00.CSV (counts 
1~50000), GW000\GW000-01.CSV (counts 
50001~100000), and GW000\GW000-02.CSV 
(counts 100001 ~ 102000).  
 Note 
*Please note that automatic file name generation 
only occurs if the FILE setting is set to NEW FILE. 
See page 92 for details. 
**Please note that the suffix will only be 
incremented if the total number of readings 
exceeds 50000. To be able to exceed 50000 
readings, either the FILE setting should be set to 
CONTINU (continuous) or the Count setting 
should be set to CONTINU (continuous). See page 
91 and 92 for details. 
Operator Mode 
 
Overview 
In the operator mode, you can choose to 
operate in Simple mode or in Advance mode, 
where various parameters can be designated by 
the user. 
Simple Mode 
This mode is the easiest operation mode and is 
almost setting free. It is the default operating 
mode. After entering this mode, the system will 
set the ‘Existing File’ setting to ‘New File,’ 
‘Count’ to ‘Continu,’ and ‘Time Mode’ to 
‘Restart’ by default. The system will then start 
to seek for the first available file name (e.g. The 
first file name will usually start from GW000, if 
GW000 doesn’t already exist). If GW000 and 
GW001 exist already, then GW002 would be the 
next available filename.