Fluke , 9 Hz thermography camera, , 80 x 60 pix bolometer matrix 4484837 User Manual

Product codes
4484837
Page of 68
 Thermal 
Imagers 
 
Menus 
 
37 
Note 
The Imager does not sense ambient or surface dew point level 
automatically. To use the low-temperature color alarm function as a 
dew point color alarm, manual determination and input of surface dew 
point temperature will yield the best results. Depending on the 
situation, the colors presented will, or will not, actually show areas with 
possible dew point condensation. 
To view the Color Alarm menu: 
1.  Push . 
2.  Push / to highlight Image
3.  Push  or  to view the menu. 
4.  Push / to highlight Color Alarm
5.  Push  or  to view the menu. 
Set High-Temperature Color Alarm 
To set a high-temperature color alarm: 
1.  From the Color Alarm menu, push / to highlight the option: Set 
High Alarm
2.  Push  to open the Color Alarm menu. 
3.  Push / to adjust the temperature setting. 
4.  Push: 
•   to set the change and go back to the live view. 
•   or  to set change and go back to the previous menu. 
•   to cancel the change and go back to the live view