Fluke , 9 Hz thermography camera, , 80 x 80 pix bolometer matrix 4484843 User Manual

Product codes
4484843
Page of 68
 Thermal 
Imagers 
 
Focus 
 
13 
Focus 
Models Ti110, Ti125, TiR110, and TiR125 have IR-OptiFlex focus. The Ti90, 
Ti95, Ti100, Ti105, and TiR105 use a large depth of field focus-free system. 
Models with IR-OptiFlex focus can operate in a focus-free mode, but also have 
the flexibility for close focus situations (<122 cm/48 inches) with a one-touch, 
fine-tune focus capability. 
The focus-free system can focus at a minimum distance of 122 cm (48 in) and 
beyond with no adjustment.  
Correct focus is important in all imaging applications. Correct focus makes sure 
that the infrared energy is correctly directed onto the pixels of the detector. 
Without the correct focus, the thermal image can be blurry and the radiometric 
data will be inaccurate. Out-of-focus infrared images are frequently unusable or 
of little value. 
To operate the IR-OptiFlex focus in the focus-free mode, align the white dot on 
the focus control with the white dot on the body of the Imager. You will also feel 
a detent at this position. See Figure 2. In this mode, in addition to correctly 
focused infrared images, the IR-Fusion must always be in proper alignment. 
 
gju04.eps 
Figure 2. IR-OptiFlex Focus 
4 ft
>4 ft