Reflected Background Temperature Compensation 
Compensation for reflected background temperature is set in the Background 
tab. Very hot objects or very cold objects can affect the apparent temperature 
and measurement accuracy of the target or object of interest, especially when 
surface emissivity is low. Adjustment of the reflected background temperature 
can make the temperature measurement better in many situations. For more 
information, see Emissivity Adjustment
To adjust the background temperature: 
1.  Push . 
2.  Push / to highlight Measurement
3.  Push  or  to view the menu. 
4.  Push / to highlight Background
5.  Push  or  to view the menu. 
6.  Push / to change the value. 
7.  Push: 
•   to set the change and go back to the live view. 
•   or  to set the change and go back to the previous menu. 
•   to cancel the change and go back to the live view. 
Note 
If the Display is set to Display All, you see the information about 
current reflected background temperature as 
BG = xx.x
.