Transmission/Transmittance Adjustment 
When you do infrared inspections through infrared-transparent windows 
(IR windows), not all of the infrared energy emitted from the objects of interest 
is transmitted through the optical material in the window. If the transmission 
percentage of the window is known, you can adjust this percentage in the 
Imager or in the SmartView
 software. Adjustment of the transmission 
correction can make the accuracy of the temperature measurement better in 
many situations. 
To adjust the transmission percentage: 
1.  Push . 
2.  Push / to highlight Measurement
3.  Push  or  to view the menu. 
4.  Push / to highlight Transmission
5.  Push  or  to view the menu. 
6.  Push / to adjust the percentage. 
7.  Push: 
•   to set the change and go back to the live view. 
•   or  to set the change and go back to the previous menu. 
•   to cancel the change and go back to the live view. 
Note 
If Display Information is set to Display All, you see the information 
about current transmission correction as 
τ 
xx
.